Retractable BSE detector for SEM for hot stage samples
Key properties
Reliable sensor
- Single crystal sensor
- Scintillator/PMT principle for ultimate image quality
- Excellent S/N ratio
- Unlimited sensor lifetime
- HV + LV + ESEM operation
- Sample temperature over 1000°C
Precies mechanics
- Motorized retraction mechanism
- Bellows-sealed high vacuum system
- Retraction 150 mm
- Compact design
- Precise alignment
- Customized connection to SEM
Examples of measurements acquired with KARMEN™
Images with courtesy Podor R, Mendonça J, Lautru J, et al. Evaluation and application of a new scintillator-based heat-resistant back-scattered electron detector during heat treatment in the scanning electron microscope. Journal of Microscopy. 2020;1-15.
Sensor heat management
- Sample temperature 1000°C
- Passive cooling by heat-pipe technology
- Temperature monitoring of the sensor
Sensor heat shielding
- Sensor and lightguide cooled and
shielded against infrared and visible
light emited by the heated sample
Sensor unit

- single crystal scintillator with optimized efficiency and minimized afterglow, exponential decay time 45 ns with 30 photons/keV
- Outer diameter 15 mm
- Inner hole diameter 3 / 2 / 1.5 or 1 mm. This dimension limits the field of view
- Exclusive CRYTUR LOW-ENERGY COATING™ sensitive down to 0.2 keV
- Sensitivity down to 1pA beam current
- Thin - allows a working distance down to 2 mm
Mechanical unit

- Dimensions 406 x 100 x 72 mm
- Bellows design for ultimate vacuum
- 150 mm motorized retraction
- +/- 2 mm fine adjustment in x, y, z axis
- 0.02 mm repeatability
- Flange adapted to customer's SEM
- Silver / black / blue / red matt surface finish available
Electronics Control Unit

- Dimensions 270 x 180 x 65 mm
- Supply 230/110 V
- 3 noise filters optimized for different scanning speeds
- 50 MHz bandwidth
- Internal source 0-1200 V for PMT HV
- Output video signal -1/+1V (-5/+5V)
- Full remote software control
- DC offset up to -20V